FDA Adverse Event Malfunction Summary report: N

CL NON-DEHP CONTINU-FLO SET10 DPM, 110", 3 CL SITES

MDR report key: 1973412 · Received January 30, 2011

Report

Report Number
6000001-2011-00504
Event Type
Malfunction
Date Received
January 30, 2011
Date of Event
January 1, 2011
Report Date
January 4, 2011
Manufacturer
BAXTER HEALTHCARE
Product Code
FPA
PMA / PMN Number
K003225
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER RETURNED ONE USED SAMPLE FOR EVALUATION. THE SAMPLE WAS VISUALLY AND FUNCTIONALLY TESTED AND THE REPORTED CONDITION WAS NOT CONFIRMED. THE SAMPLE PASSED VISUAL INSPECTION, SLIT VISUALIZATION TESTING, CLEAR AIR PASSAGE TESTING AT 0.25 PSI FOR 30 SECONDS, REFEREE TESTING OF THE CLEARLINK Y-SITES AT 0.25 PSI FOR 30 SECONDS, AND FLUID PATH OCCLUSION TESTING. THEREFORE, AN ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED. A BATCH REVIEW COULD NOT BE PERFORMED AS THE LOT NUMBER IS UNKNOWN. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS REPORTED TO BE AVAILABLE, BUT HAS NOT BEEN RECEIVED FOR EVALUATION. IF ADDITIONAL INFORMATION BECOMES AVAILABLE OR THE SAMPLE IS RECEIVED, A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CORPORATE PRODUCT SURVEILLANCE THAT THEY ARE HAVING DIFFICULTY ESTABLISHING FLOW AT THE UPPER Y-SITE OF THE CLEARLINK NON-DEHP CONTINU-FLO SOLUTION SET AFTER CONNECTING AN UNKNOWN BAXTER SECONDARY SET. THEY DON'T NORMALLY TAP AND INVERT THE BACKCHECK VALVE. THE CONDITION OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO OTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CL NON-DEHP CONTINU-FLO SET10 DPM, 110", 3 CL SITES SET, ADMINISTRATION, INTRAVASCULAR FPA BAXTER HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN BAXTER SECONDARY SET