PRECISION XTRA
Report
- Report Number
- 2954323-2011-01263
- Event Type
- Malfunction
- Date Received
- January 28, 2011
- Date of Event
- November 15, 2010
- Report Date
- January 28, 2011
- Manufacturer
- ABBOTT DIABETES CARE INC.
- Product Code
- NBW
- Removal / Correction Number
- ADC FA1197-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
RETAIN SAMPLES OF PRECISION TEST STRIP LOTS MANUFACTURED WITH HOT-MELT GLUE BATCHES EXHIBITED LOWER THAN EXPECTED READINGS ON CONTINUOUS STORAGE AFTER NINE MONTHS AT 30 DEGREES CELSIUS DURING ROUTINE STABILITY PERFORMANCE TESTING. THIS STABILITY ISSUE COULD LEAD TO THE GENERATION OF INCORRECTLY DEPRESSED BLOOD GLUCOSE RESULTS AND THESE ERRONEOUS RESULTS MAY BE IN THE "C", "D" OR "E" ZONES OF PARKES ERROR GRID, AND AS SUCH, HAVE THE POTENTIAL TO BE CLINICALLY SIGNIFICANT. THE PRIMARY CAUSE HAS BEEN IDENTIFIED TO BE BATCH TO BATCH GLUE VARIABILITY. THE FDA HAS BEEN INFORMED OF THE FIELD ACTION PER 21CFR806 (FIELD ACTION REFERENCE NUMBER 2954323-12/22/10-001-R). ALL AFFECTED CONSIGNEES WERE NOTIFIED BY LETTER BEGINNING DECEMBER 22, 2010.
CUSTOMER REPORTED THAT STARTING IN (B)(6) 2010 HE BEGAN TO EXPERIENCE BLURRED VISION. HE FURTHER REPORTED ADDITIONALLY EXPERIENCING "BLOODY NOSES AND RECENTLY HEADACHES". NO THIRD-PARTY EMERGENT INTERVENTION WAS REPORTED. CUSTOMER SELF-TREATED BY DRINKING WATER AND TAKING ASPIRIN. CUSTOMER FURTHER REPORTED RECEIVING THE FOLLOWING READINGS WHICH WERE LOWER THAN HE FELT: 98 (B)(6) 2010, 80 (B)(6) 2010, 89 (B)(6) 2010, 89 (B)(6) 2010, 99 (B)(6) 2010, 118 (B)(6) 2010 AND 132 (B)(6) 2010. IT WAS ADDITIONALLY REPORTED THE CUSTOMER HAD BEEN USING TEST STRIPS RELATED TO AN ON-GOING FIELD ACTION FOR ABBOTT'S PRECISION FAMILY TEST STRIPS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THESE EVENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRECISION XTRA | BLOOD GLUCOSE MONITORING SYSTEM | NBW | ABBOTT DIABETES CARE INC. | 45001A788 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |