OT PING METER
Report
- Report Number
- 2939301-2011-01012
- Event Type
- Malfunction
- Date Received
- January 28, 2011
- Report Date
- January 24, 2011
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 2/22/2011 WITH THE FOLLOWING FINDINGS: THE METER HAS PASSED TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K)# IS K082590.
THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE OK BUTTON IS STUCK/STICKING AND WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | 3066958 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |