FDA Adverse Event Malfunction Summary report: N

CE INFUSOR, LV 5 ML/H,12 PACK

MDR report key: 1973260 · Received January 28, 2011

Report

Report Number
6000001-2011-00517
Event Type
Malfunction
Date Received
January 28, 2011
Date of Event
January 3, 2011
Report Date
January 4, 2011
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: ONE SAMPLE WAS RECEIVED BY BAXTER FOR EVALUATION. THE REPORTED CONDITION OF "RUPTURED RESERVOIR" WAS CONFIRMED. THIS IS A SINGLE USE DEVICE AND WILL NOT BE REPAIRED. THE ROOT CAUSE WILL BE IDENTIFIED/ADDRESSED THROUGH THE CAPA INVESTIGATION, (B)(4). A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL NARRATIVE: THE DEVICE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION AND/OR SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER CHINA THAT THE RESERVOIR OF ONE (1) CE INFUSOR HAD RUPTURED DURING FILLING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CE INFUSOR, LV 5 ML/H,12 PACK PUMP, INFUSION, ELASTOMERIC MEB 09N004

Patients

Seq Age Sex Outcome Treatment
1