FDA Adverse Event Malfunction Summary report: N

CE INFUSOR, LV 5 ML/H,12 PACK

MDR report key: 1973248 · Received January 28, 2011

Report

Report Number
6000001-2011-00496
Event Type
Malfunction
Date Received
January 28, 2011
Date of Event
March 25, 2010
Report Date
January 4, 2011
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL NARRATIVE: THE DEVICE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION AND/OR SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (U.S.); THEREFORE, IT DOES NOT CONTAIN A U.S. 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE U.S.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. A VISUAL EXAMINATION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION COULD NOT CONFIRM THE REPORTED CONDITION OF BACKFLOW FROM THE FILLING PORT. THE ROOT CAUSE WAS DETERMINED TO BE A GLASS FRAGMENT TRAPPED BETWEEN THE CHECKBAND AND THE STRESSMEMBER. THIS DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(6) THAT ONE (1) CE INFUSOR DEVICE WAS OBSERVED LEAKING DURING FILLING. IT IS UNKNOWN WHERE THE LEAK IS LOCATED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CE INFUSOR, LV 5 ML/H,12 PACK PUMP, INFUSION, ELASTOMERIC MEB 09B007

Patients

Seq Age Sex Outcome Treatment
1