FDA Adverse Event Death Summary report: N

CAPSURE SP NOVUS

MDR report key: 1973177 · Received January 28, 2011

Report

Report Number
2649622-2011-00146
Event Type
Death
Date Received
January 28, 2011
Date of Event
January 4, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P830061/S27
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) OUTER INSULATION BREACHED AND COSMETIC ESC AND COSMETIC MIO. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO MANFACTURER FOLLOWING THE PATIENT'S DEATH, ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO MANUFACTURER FOLLOWING THE PATIENT'S DEATH, ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. FOLLOW UP NOTED DISCHARGE SUMMARY FROM HOSPITAL REPORTS PATIENT WAS ADMITTED TO THE HOSPITAL 13 DAYS PRIOR TO PATIENT DEATH. CAUSE OF DEATH WAS CONGESTIVE HEART FAILURE, CHRONIC RENAL FAILURE, PNEUMONIA, RESPIRATORY FAILURE, SEPSIS. PATIENT ALSO HAD CHRONIC OBSTRUCTIVE PULMONARY DISEASE WITH A TRACHEOTOMY AND WAS POSITIVE FOR PSEUDOMONAS INFECTION. THE DISCHARGE SUMMARY DID NOT STATE THAT THE DEVICE AND/OR LEADS WERE IMPLICATED IN THE PATIENT'S DEATH. ALSO THERE WAS NO MENTION OF ANY PROBLEMS WITH THE DEVICE OR LEADS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SP NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4592 ASKU

Patients

Seq Age Sex Outcome Treatment
1 85 YR Death 4092 IMPLANTABLE PACING LEAD| 4092 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR| (B)(4) IMPLANTABLE PULSE GENERATOR