FDA Adverse Event Malfunction Summary report: N

REGIONAL ANALGESIA INFUSOR W/PATIENT CONTROL MODULE 5.0-7.0

MDR report key: 1973126 · Received January 28, 2011

Report

Report Number
6000001-2011-00478
Event Type
Malfunction
Date Received
January 28, 2011
Date of Event
December 15, 2010
Report Date
January 3, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K051253
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL NARRATIVE: THE DEVICE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION AND/OR SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: ONE SAMPLE WAS RECEIVED BY BAXTER FOR EVALUATION. THE REPORTED CONDITION OF "LEAKING AT THE BLUE WINGED CAP" WAS CONFIRMED. THIS IS A SINGLE USE DEVICE AND WILL NOT BE REPAIRED. NO OTHER OBSERVATION WAS FOUND. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT ONE (1) REGIONAL ANALGESIA INFUSOR W/PATIENT CONTROL MODULE WAS OBSERVED LEAKING AT THE BLUE WINGED CAP AFTER FILLING. THE DEVICE WAS FILLED WITH NAROPIN. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 4 OF 7 WITH THE SAME REPORTED PROBLEM FROM THIS FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REGIONAL ANALGESIA INFUSOR W/PATIENT CONTROL MODULE 5.0-7.0 PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 10E065

Patients

Seq Age Sex Outcome Treatment
1 NAROPIN