FDA Adverse Event Other Summary report: N

KRYPTONITE BONE CEMENT

MDR report key: 1973099 · Received January 21, 2011

Report

Report Number
1226001-2011-00001
Event Type
Other
Date Received
January 21, 2011
Date of Event
December 8, 2010
Report Date
December 14, 2010
Manufacturer
DOCTOR'S RESEARCH GROUP, INC.
Product Code
GXP
PMA / PMN Number
K091382
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL LOT #: 12182009. ADDITIONAL MFR DATE: 12/18/2009. THE SPEED OF A CHEMICAL REACTION IS RELATED TO TEMPERATURE AND THE AUTOCATALYTIC AFFECT OF AN EXOTHERMIC REACTION. SIMILAR TO TEMPERATURE INFLUENCES; MATERIAL VOLUMES AND THE SHAPE OF THE MATERIAL WHILE CURING (SHEET VS. BALL) ALSO CONTRIBUTE TO THE INCREASE OF AN EXOTHERMIC REACTION. FURTHER TO THESE AUTOCATALYTIC AND VOLUMETRIC AFFECTS, ALSO NOTE THAT SURGICAL SITE (I.E., A WET ENVIRONMENT, ETC.) ALSO CONTRIBUTES TO SLOWING A REACTION. THE REACTION WOULD THEREFORE INCREASE FROM A 5CC KIT (SLOWER REACTION) TO A 10CC KIT (FASTER REACTION). THE OBSERVED MATERIAL CONSISTENCY, IS LIKELY THE RESULT OF THE EXCESSIVE CEREBRAL FLUID NOTED BY THE SURGEON DURING THE INITIAL SURGERY. THE PRODUCT INSTRUCTIONS CLEARLY STATE SURGICAL WOUND AREA MUST BE DRY FOR BEST PERFORMANCE OF THE PRODUCT. NO DOCUMENTATION WITHIN THE DHR WAS FOUND THAT WOULD INDICATE A NONCONFORMANCE TO SPECIFICATION THAT COULD HAVE CONTRIBUTED TO THIS EVENT. IF ADDITIONAL INFO BECOMES AVAILABLE, IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

A PT UNDERWENT A CRANIAL FRACTURE, POSTERIOR ORBIT-INTRACRANIAL PROCEDURE WHERE KRYPTONITE BONE CEMENT WAS USED. A DURAL TEAR WAS OBSERVED WITH A SIGNIFICANT CSF LEAK. THE PERICARDIUM WAS PULLED OVER AND DURA SEAL SPRAYED ON SITE. DURA SEAL WAS NOT SPRAYED DIRECTLY ON THE KRYPTONITE. IT WAS NOTED, HOWEVER, THAT SOME MATERIAL MAY HAVE COME IN CONTACT WITH KRYPTONITE. UNUSED MATERIAL WAS PLACED ON THE SURGICAL TABLE FOR LATER USE IN FILLING THE BURR HOLES. THE KRYPTONITE MATERIAL WAS LEFT IN THE TRAY AND PLACED IN AN ICE BATH TO SLOW THE POLYMERIZATION PROCESS. WHEN THE SURGEON RETRIEVED THAT MATERIAL TO FILL THE BURR HOLES, THE MATERIAL HAD POLYMERIZED (HARDENED). THE SURGEON USED A SECOND KIT TO FILL THE BURR HOLES. A REVISION WAS PERFORMED THE NEXT DAY DUE TO INCREASED CRANIAL PRESSURE, LIKELY DUE TO THE CONTINUED CSF LEAK. WHILE REVISING THE SITE, IT WAS OBSERVED THAT THE KRYPTONITE BONE CEMENT WAS NOT AS HARD AS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KRYPTONITE BONE CEMENT BONE CEMENT GXP DOCTOR'S RESEARCH GROUP, INC. KRYP-US-Z-05 11062010

Patients

Seq Age Sex Outcome Treatment
1 UNK