FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1973083 · Received January 28, 2011

Report

Report Number
1423500-2011-01217
Event Type
Malfunction
Date Received
January 28, 2011
Date of Event
December 3, 2010
Report Date
January 5, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE DEVICE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE SYSTEM ERROR (SE) 2240 WAS IDENTIFIED DURING A REVIEW OF A RETURNED HOMECHOICE (HC) DEVICE WITH OCCURRENCE ON (B)(6) 2010; THERE WAS NO REPORT FOR THIS ALARM CALLED IN BY THE CUSTOMER. THERE IS NO EVIDENCE THAT PROBLEMS WITH THE HC CONTRIBUTED TO SE 2240. A ROOT CAUSE WAS NOT DETERMINED. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

A SYSTEM ERROR (SE) 2240 ALARM WAS IDENTIFIED IN THE LOGS DURING EVALUATION OF A HOMECHOICE (HC) DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 34 YR HOMECHOICE CYCLER