FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1973051
·
Received January 21, 2011
Report
- Report Number
- 2531779-2011-00399
- Event Type
- Malfunction
- Date Received
- January 21, 2011
- Report Date
- December 26, 2010
- Manufacturer
- ANIMAS CORP
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED, THE CONTRAST/UP/DOWN KEYPAD BUTTON WAS HYPERSENSITIVE TO USER INPUTS DURING TESTING, IT WAS OBSERVED THAT THE OK KEY CONTACT WAS INVERTED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE UP/DOWN/OK/KEYPAD BUTTONS INTERMITTENTLY UNRESPONSIVE WHEN PRESSED. THE REPORTER CLAIMED THAT THE KEYPAD IS STILL INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |