FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1973051 · Received January 21, 2011

Report

Report Number
2531779-2011-00399
Event Type
Malfunction
Date Received
January 21, 2011
Report Date
December 26, 2010
Manufacturer
ANIMAS CORP
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED, THE CONTRAST/UP/DOWN KEYPAD BUTTON WAS HYPERSENSITIVE TO USER INPUTS DURING TESTING, IT WAS OBSERVED THAT THE OK KEY CONTACT WAS INVERTED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE UP/DOWN/OK/KEYPAD BUTTONS INTERMITTENTLY UNRESPONSIVE WHEN PRESSED. THE REPORTER CLAIMED THAT THE KEYPAD IS STILL INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1