FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1973017 · Received January 20, 2011

Report

Report Number
2531779-2011-00383
Event Type
Malfunction
Date Received
January 20, 2011
Report Date
December 22, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR INVESTIGATION. EVAL REVEALED THE FOLLOWING: THE BATTERY WAS NOT RETURNED WITH THE PUMP, THE BLACK BOX DOWNLOAD CONFIRMED A BATTERY VOLTAGE DROP FROM 163 VDC AT 19:13 ON (B)(6) 2010 TO 128 VDC AT 4:06 AM ON (B)(6) 2010, THE BATTERY COMPARTMENT APPEARED TO BE INTACT, NO EVIDENCE OF MOISTURE INGRESS, THE PUMP POWERED ON NORMALLY AND WAS EXERCISED FOR 9 HRS WITH NO OVERHEATING OR ALARMS, THE PUMP ELECTRICAL CURRENT TESTS WERE WITHIN SPECIFICATIONS, NO DEFECTS WERE FOUND WITH THE POWER FLEX, AND BATTERY CONNECTIONS AND INTERNAL COMPONENTS. IN CONCLUSION, THE ALLEGED ISSUE COULD NOT BE DUPLICATED DURING INVESTIGATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP DISPLAYED THE LOW BATTERY MESSAGE AND THAT THE PUMP WAS HOT. THE PT STARTED USING THE PUMP 5 DAYS AGO AND WAS USING THE ULTIMATE LITHIUM BATTERY THAT CAME UP WITH THE PUMP. THE RPTR DENIED ANY SIGNS OF CORROSION OR ANYTHING ABNORMAL IN THE BATTERY COMPARTMENT. THERE WERE NO REPORTED INJURIES AS A RESULT OF THE HOT PUMP. THE LITHIUM BATTERY WAS REPLACED WITH AN ALKALINE BATTERY: THE PT'S MOM STATED THAT THE PUMP APPEARS TO BE FUNCTIONING FINE. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1