ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
Report
- Report Number
- 2531779-2011-00383
- Event Type
- Malfunction
- Date Received
- January 20, 2011
- Report Date
- December 22, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR INVESTIGATION. EVAL REVEALED THE FOLLOWING: THE BATTERY WAS NOT RETURNED WITH THE PUMP, THE BLACK BOX DOWNLOAD CONFIRMED A BATTERY VOLTAGE DROP FROM 163 VDC AT 19:13 ON (B)(6) 2010 TO 128 VDC AT 4:06 AM ON (B)(6) 2010, THE BATTERY COMPARTMENT APPEARED TO BE INTACT, NO EVIDENCE OF MOISTURE INGRESS, THE PUMP POWERED ON NORMALLY AND WAS EXERCISED FOR 9 HRS WITH NO OVERHEATING OR ALARMS, THE PUMP ELECTRICAL CURRENT TESTS WERE WITHIN SPECIFICATIONS, NO DEFECTS WERE FOUND WITH THE POWER FLEX, AND BATTERY CONNECTIONS AND INTERNAL COMPONENTS. IN CONCLUSION, THE ALLEGED ISSUE COULD NOT BE DUPLICATED DURING INVESTIGATION.
IT WAS REPORTED THAT THE PUMP DISPLAYED THE LOW BATTERY MESSAGE AND THAT THE PUMP WAS HOT. THE PT STARTED USING THE PUMP 5 DAYS AGO AND WAS USING THE ULTIMATE LITHIUM BATTERY THAT CAME UP WITH THE PUMP. THE RPTR DENIED ANY SIGNS OF CORROSION OR ANYTHING ABNORMAL IN THE BATTERY COMPARTMENT. THERE WERE NO REPORTED INJURIES AS A RESULT OF THE HOT PUMP. THE LITHIUM BATTERY WAS REPLACED WITH AN ALKALINE BATTERY: THE PT'S MOM STATED THAT THE PUMP APPEARS TO BE FUNCTIONING FINE. THE PUMP WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |