FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1973004
·
Received January 20, 2011
Report
- Report Number
- 2531779-2011-00359
- Event Type
- Malfunction
- Date Received
- January 20, 2011
- Report Date
- December 22, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED, THE CONTRAST/UP/DOWN/OK KEYPAD BUTTONS INTERMITTENTLY RESPONDING TO USER INPUTS DURING TESTING, IT WAS OBSERVED THAT THE CONTRAST/UP/DOWN/OK KEY CONTACTS BECAME INVERTED WHEN PRESSED AND DID NOT ALWAYS SPRING BACK AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE UP, DOWN AND OK KEYPAD BUTTONS INTERMITTENTLY NOT RESPONDING WHEN PRESSED. IT WAS REPORTED THAT THE KEYPAD IS STILL INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANIMAS 2020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |