FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1972927 · Received January 19, 2011

Report

Report Number
2531779-2011-00339
Event Type
Malfunction
Date Received
January 19, 2011
Report Date
December 22, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVAL. IF THE DEVICE IS RETURNED, AN EVAL SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILLED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. THE REPORT IS UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (B)(6) 2011 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS OBSERVED TO BE TORN AT THE ARROW BUTTONS. THE SCREEN SCROLLED WITHOUT ANY BUTTON PRESSES. THE UP, DOWN, AND OK BUTTONS WERE NOT RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND THE DOWN ARROW BUTTON CONTACT WAS FOUND TO BE MISALIGNED; THE DOWN ARROW BUTTON WAS RESEATED AND THE BUTTONS RESPONDED APPROPRIATELY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE KEYPAD BUTTONS REPORTEDLY NOT RESPONDING WHEN PRESSED. THE REPORTER CLAIMED NO SIGNS OF PEELING OR DAMAGE TO THE KEYPAD. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020

Patients

Seq Age Sex Outcome Treatment
1