FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1972926 · Received January 19, 2011

Report

Report Number
2531779-2011-00338
Event Type
Malfunction
Date Received
January 19, 2011
Report Date
December 22, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED, THE DOWN KEYPAD BUTTON DID NOT RESPOND TO USER INPUTS DURING TESTING, THE DOWN KEYPAD BUTTON ALSO DID NOT SPRINGBACK WHEN PRESSED, IT WAS OBSERVED THAT THE DOWN KEY CONTACT WAS INVERTED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE OK, UP AND DOWN ARROW KEYPAD BUTTONS INTERMITTENTLY NOT RESPONDING WHEN PRESSED. THE REPORTER CLAIMED THAT THE BUTTONS DO NOT CLICK BACK AND SPRING AS USUAL; HOWEVER, KEYPAD INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1