ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
Report
- Report Number
- 2531779-2011-00338
- Event Type
- Malfunction
- Date Received
- January 19, 2011
- Report Date
- December 22, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED, THE DOWN KEYPAD BUTTON DID NOT RESPOND TO USER INPUTS DURING TESTING, THE DOWN KEYPAD BUTTON ALSO DID NOT SPRINGBACK WHEN PRESSED, IT WAS OBSERVED THAT THE DOWN KEY CONTACT WAS INVERTED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE OK, UP AND DOWN ARROW KEYPAD BUTTONS INTERMITTENTLY NOT RESPONDING WHEN PRESSED. THE REPORTER CLAIMED THAT THE BUTTONS DO NOT CLICK BACK AND SPRING AS USUAL; HOWEVER, KEYPAD INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |