FDA Adverse Event Malfunction Summary report: N

ENDOPATH** ETS

MDR report key: 1972901 · Received January 28, 2011

Report

Report Number
3005075853-2011-00339
Event Type
Malfunction
Date Received
January 28, 2011
Date of Event
December 16, 2010
Report Date
December 28, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE SCW45 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED FULLY LOADED WITH STAPLES. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. WHILE NO CONCLUSION COULD BE REACHED AS TO WHAT CAUSED THE FIRING MECHANISM TO FAIL, IT IS POSSIBLE THAT THE DEVICE WAS FIRED ON TISSUE THICKER THAN INDICATED OR THROUGH A LOCKED CARTRIDGE IN PREVIOUS FIRINGS. WHEN EITHER OR BOTH OF THESE SCENARIOS OCCUR THE COMPONENTS IN THE FIRING MECHANISM CAN BE DAMAGED. IT SHOULD BE NOTED THAT A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; (B)(4). A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE LOADED WITH THE ORIGINAL CARTRIDGE COULD BE FIRED PROPERLY AT THE FIRST FIRING BUT WHEN THE DEVICE WAS FIRED ON THE SECOND FIRING THE KNIFE DID NOT MOVE FORWARD AND THE STAPLES WERE NOT DEPLOYED AT ALL. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** ETS STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK E4LR23

Patients

Seq Age Sex Outcome Treatment
1