FDA Adverse Event Injury Summary report: N

INTERNA UNIVERSAL PLATFORM IMPLANT CA

MDR report key: 19728736 · Received July 12, 2024

Report

Report Number
3004417597-2024-00023
Event Type
Injury
Date Received
July 12, 2024
Date of Event
December 2, 2020
Manufacturer
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
Product Code
DZE
UDI-DI
08435389812128
PMA / PMN Number
K151391
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE BATCH DOCUMENTATION IS CHECKED AND NO ANOMALIES ARE DETECTED. ANY MANUFACTURING FAULT IS RULED OUT. AFTER EXAMINATION OF THE IMPLANT, IT CAN BE VERIFIED THAT THE IMPLANT DOES NOT SHOW SIGNS OF MISUSE OR SIGNS OF MALFUNCTION THAT COULD HAVE CONTRIBUTED TO THE EVENT.

Description of Event or Problem · 0

THE CLINICIAN INDICATES THAT HE PLACED THE IMPLANT ON (B)(6) 2020 AND ONE MONTH LATER IT WAS NOTED THAT THE IMPLANT HAD NOT OSSEOINTEGRATED. PATIENT WITH A BONE QUALITY BETWEEN II AND III. IN THE EVENT THE PATIENT EXPERIENCED: INFECTION, BONE LOSS, PAIN AND INFLAMMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2386504 INTERNA UNIVERSAL PLATFORM IMPLANT CA DZE B.T.I. BIOTECHNOLOGY INSTITUTE S.L. IIPUCA3711 BO01761.Z 08435389812128

Patients

Seq Age Sex Outcome Treatment
1 72 YR Male Required Intervention