ETS FLEX
Report
- Report Number
- 3005075853-2011-00334
- Event Type
- Malfunction
- Date Received
- January 28, 2011
- Date of Event
- January 2, 2011
- Report Date
- January 3, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). INSTRUMENT B: PINION AXLE SUPPORT. THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE (A) WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS NO CARTRIDGE WAS RECEIVED FOR ANALYSIS. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN (B)(4), AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RETURNED WITH THE FIRING MECHANISM DAMAGED AND WITH NO RELOAD PRESENT. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE LEFT SHROUD PINION AXLE SUPPORT WAS FOUND DAMAGED. WHILE NO CONCLUSION COULD BE REACH WHAT CAUSE THE FIRING MECHANISM TO FAIL, IT IS POSSIBLE THAT THE DEVICE WAS FIRED ON TISSUE THICKER THAN INDICATED OR THROUGH A LOCKED CARTRIDGE. WHEN EITHER OR BOTH OF THESE SCENARIOS OCCUR THE COMPONENTS IN THE FIRING MECHANISM CAN BE DAMAGED. IT SHOULD BE NOTED THAT A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; (B)(4). NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT, AS THERE WAS NO RELOAD RETURNED FOR ANALYSIS.
THE PROCEDURE WAS A LAPAROSCOPIC APPENDECTOMY. HALF THE STAPLES CAME DOWN AND THE BLADE DID NOT CUT. THERE WERE NO PATIENT CONSEQUENCES. ANOTHER DEVICE WAS USED AND THE SAME THING HAPPENED HALF THE STAPLES CAME DOWN AND THE BLADE DID NOT CUT. THE SECOND DEVICE WAS RELOADED AND THE DEVICE FINALLY WORKED. THE CALLER MENTIONED THAT SHE IS NOT SURE IF THE ISSUE WAS USER ERROR. THE SURGEON'S NAME IS DR. (B)(6). THE 2 DEVICES WILL BE RETURNED HOWEVER AN UNKNOWN NUMBER OF RELOADS WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETS FLEX | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | G4U56E |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |