FDA Adverse Event Other Summary report: N

TRIMA ACCEL (PLT)

MDR report key: 1972811 · Received January 14, 2011

Report

Report Number
1722028-2010-00154
Event Type
Other
Date Received
January 14, 2011
Date of Event
February 11, 2010
Report Date
January 14, 2011
Manufacturer
CARIDIANBCT
Product Code
GKT
PMA / PMN Number
BK030083
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION: THE ACDA SOLUTIONS USED WERE FROM TWO DIFFERENT LOTS. THE TWO INCIDENTS THAT OCCURRED DURING A TRIMA ACCEL COLLECTION USED TRIMA DISPOSABLE KITS FROM THE SAME LOT NUMBER 12R1115. SYNCOPE EPISODES HAVE SIMILAR CHARACTERISTICS, BUT CAN BE CAUSED BY MANY DIFFERENT TYPES OF STIMULI. ROOT CAUSE: THESE DISPOSABLE SETS WERE UNAVAILABLE FOR SPECIFIC ROOT CAUSE ANALYSIS. THE PHYSICIAN POINTED TO DEHYDRATION AS THE ROOT CAUSE OF BOTH INCIDENTS. RISK ANALYSIS: THE TWO INDIVIDUALS EXPERIENCED A TYPE OF REACTION THAT IS LISTED IN THE TRIMA ACCEL AUTOMATED BLOOD COLLECTION OPERATOR'S MANUAL (TRIMA, 2009). THE RATE OF OCCURRENCE OF ALL REPORTED PT REACTIONS ON TRIMA DISPOSABLE KITS MFG IN 2009 WAS LESS THAN 1 IN 100,000. CONCLUSION: THE DEVICE OPERATED AS INTENDED. HYPOCALCEMIC REACTIONS ARE A KNOWN SIDE EFFECT OF APHERESIS PROCEDURES.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS A RESULT OF CHANGES TO OUR MDR EVAL PROCESS. WE HAVE CHANGED OUR PROCESS TO BETTER ALIGN WITH CURRENT AGENCY POLICY. DONOR PALE, LIGHTHEADED, HOT AND DIAPHORETIC. LOC FOR 30-45 SEC WITH TETANY. AMMONIA INHALANT USED AND DONOR PLACED IN SUPINE POSITION FLUIDS GIVEN, HEAD LOWERED, KNEES ELEVATED, COUGH ENCOURAGED, COLD PACK APPLIED, AND INSTRUCTIONS GIVEN. PROLONGED RECOVERY NOTED, OVER 1 HR. DISCHARGED UNDER CARE OF WIFE. STAFF CALLED LATER IN THE DAY, DONOR FEELING GOOD. FOLLOW UP CALL NEXT DAY, DONOR FINE WITH NO COMPLAINTS. THIS REPORT IS BEING FILED DUE TO MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRIMA ACCEL (PLT) SEPARATOR, AUTOMATED, BLOOD CELL AND PL GKT CARIDIANBCT 12R1115

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention