PENTAX
Report
- Report Number
- 9610877-2024-55334
- Event Type
- Malfunction
- Date Received
- July 12, 2024
- Date of Event
- July 5, 2024
- Report Date
- July 12, 2024
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EQL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS DEVICE IS CLASS1 PRODUCT SO THAT 510K# IS BLANK. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD DRIVE PCB FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CCD DRIVE PCB. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE REMOTE CONTROL BUTTONS PERFORATED, THE CCD DRIVE PCB CORRODED, THE LIGHT GUIDE CABLE COLLAPSE, AND THE CCD MODULE WITH DRIVE PCB DISTORTED IMAGE; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341344 | PENTAX | VIDEO LARYNGOSTROBOSCOPE (SLIM) | EQL | HOYA CORPORATION PENTAX TOKYO OFFICE | VLS-1070STK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |