PENTAX
Report
- Report Number
- 9610877-2024-55344
- Event Type
- Malfunction
- Date Received
- July 12, 2024
- Date of Event
- July 4, 2024
- Report Date
- July 12, 2024
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- CAL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THIS DEVICE IS NOT DISTRIBUTED IN US SO THAT 510K# IS BLANK. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CFB (COHERENT FIBER BUNDLE) FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CFB (COHERENT FIBER BUNDLE). IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE CONTROL BODY FLUID DAMAGE, THE SIDE BODY COVER BROKEN, AND THE INSERTION FLEXIBLE TUBE PERFORATED; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT "HR-RPT-0586(IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 236379 | PENTAX | FIBEROPTICINTUBATIONSCOPE 2.4TP RO PLS | CAL | HOYA CORPORATION PENTAX TOKYO OFFICE | FP-7RBS2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |