FDA Adverse Event
Malfunction
Summary report: N
STRAIGHT MEDIUM ELITE ATTACHMENT
MDR report key: 1972750
·
Received January 21, 2011
Report
- Report Number
- 1811755-2011-00206
- Event Type
- Malfunction
- Date Received
- January 21, 2011
- Date of Event
- December 9, 2010
- Report Date
- December 9, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DRILL WAS RETURNED TO THE MFR FOR AN UNRELATED ISSUE AND UPON EVAL, AN OVERHEATING CONDITION WAS FOUND. INTERNAL COMPONENTS WERE EVALUATED AND A DAMAGED BEARING WAS DISCOVERED. DAMAGED BEARINGS CAN RESULT IN HEAT BEING GENERATED, DUE TO EXCESSIVE FRICTION BETWEEN ROTATING COMPONENTS. THE DRILL ATTACHMENT COULD NOT BE REPAIRED, AND THEREFORE, WAS NOT RETURNED TO THE USER FACILITY.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MFR, THE DRILL ATTACHMENT HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STRAIGHT MEDIUM ELITE ATTACHMENT | DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INC | HBE | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |