FDA Adverse Event Malfunction Summary report: N

VARISOFT

MDR report key: 19727149 · Received July 11, 2024

Report

Report Number
8021545-2024-02040
Event Type
Malfunction
Date Received
July 11, 2024
Date of Event
June 3, 2024
Report Date
November 12, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018723
PMA / PMN Number
K162812
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE INITIAL MDR WITH MANUFACTURING REPORT NUMBER, WAS SUBMITTED ON 11-JUL-2024. UPON COMPLETION OF THE INVESTIGATION, THE MANUFACTURING DATE WAS UPDATED AS 07-DEC-2023. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. PER REVISION 21 OF PROCEDURE (B)(4), A CHILD INVESTIGATION IS NOT REQUIRED TO DOCUMENT THE DEVICE HISTORY RECORD (DHR) REVIEW FOR A TYPE 2 REPORTABLE COMPLAINT. HOWEVER, THE CHILD WAS CREATED TO ALLOW THE PARENT RECORD TO ADVANCE TO REVIEW AND SUMMARY. SINCE IT WAS CREATED, THE DHR REVIEW WAS DOCUMENTED WITHIN THE CHILD. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6002801, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DHR REVIEW: THE LOT 6002801 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 16 AND PACKAGED IN THE MACHINE MULTIVAC 10 ON 07-DEC-2023, WITH A TOTAL OF (B)(4) UNITS. TUBE GLUING THE SUB-ASSEMBLY, TUBE GLUING OF THE LOT 3H01820 WAS MANUFACTURED ACCORDING TO THE WI VERSION 55 AND MANUFACTURED IN THE MACHINE SC06, ON 14-AUG-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT A NON-CONFORMANCES (NC) 1694785 WAS RAISED DUE TO DIFFERENCE BETWEEN INTERNAL DIMENSION. THIS NC IS NOT RELATED TO CLAIMED MALFUNCTION. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: ONE NONCONFORMANCE (NC) WAS RAISED AND FOUND UNRELATED TO THE REPORTED MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4) EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED INFUSIONS SET TUBING DISCONNECTED FROM T:LOCK EVENT ON 03-JUN-2024. INFUSION SET HAS BEEN USED FOR 3 DAYS. CUSTOMER REPLACED INFUSION SET AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303233 VARISOFT UNO COMFORT SHORT 80/13 TCAP 10PK INT FPA UNOMEDICAL A/S 1002828 6002801 05705244018723

Patients

Seq Age Sex Outcome Treatment
1 70 YR Female