CARDIOVASCULAR PROCEDURE KIT
Report
- Report Number
- 1212122-2011-00003
- Event Type
- Malfunction
- Date Received
- January 20, 2011
- Date of Event
- December 28, 2010
- Report Date
- December 28, 2010
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
- Product Code
- OEZ
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
TERUMO DID RECEIVE THE ACTUAL DEVICE AND THE COMPLAINT WAS NOT CONFIRMED. PRESSURE TESTING WAS CONDUCTED ACROSS THE DOME MEMBRANE AND LUER CONNECTOR, AND NO LEAKS OR PRESSURE DROPS WERE DETECTED. TERUMO WILL MONITOR FOR FUTURE ISSUES. THE COMPLAINT WILL BE INCLUDED IN ASSOCIATE AWARENESS TRAINING. THE DEVICE HISTORY RECORD DID NOT INDICATE ANY PRODUCTION RELATED PROBLEMS. ALL AVAILABLE INFO HAS BEEN PLACED ON FILE IN QUALITY MANAGEMENT FOR APPROPRIATE TRACKING, TRENDING, AND F/U. (B)(4).
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS SURGERY, DURING PRIME, THE ISOLATOR DOME LEAKS AND DOES NOT HOLD PRESSURE. THERE WERE 5 OCCURRENCES OF THIS EVENT. THE PRODUCT WAS CHANGED OUT, THERE WAS NO BLOOD LOSS, AND SURGERY WAS COMPLETED SUCCESSFULLY. THE EVENT DID NOT CAUSE DELAY IN SURGICAL PROCEDURE. THERE WAS NO PT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARDIOVASCULAR PROCEDURE KIT | CONVENIENCE TUBING PACK | OEZ | TERUMO CARDIOVASCULAR SYSTEMS CORPORATION | MK30T |