FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 1972651 · Received January 28, 2011

Report

Report Number
2032227-2011-00279
Event Type
Injury
Date Received
January 28, 2011
Date of Event
January 17, 2011
Report Date
January 17, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED AFTER BEING INVOLVED IN A CAR ACCIDENT. THE CUSTOMER WAS WEARING THE INSULIN PUMP AT THE TIME OF THE ACCIDENT, BUT WAS NOT EXPERIENCING HIGH OR LOW BLOOD GLUCOSE LEVELS. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP PASSED THE SELF TEST. FOUND LOW RESERVOIR AND BATTERY OUT LIMIT ALARMS IN THE ALARM HISTORY. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC MINIMED MMT-722LNAS

Patients

Seq Age Sex Outcome Treatment
1 22 YR Hospitalization