FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 1972646 · Received January 20, 2011

Report

Report Number
3004464228-2011-00025
Event Type
Malfunction
Date Received
January 20, 2011
Date of Event
December 22, 2010
Report Date
December 22, 2010
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION OF THE RETURNED POD FOUND EVIDENCE OF AN INTERNAL FLUID LEAK. DISCOLORATION WAS SEEN INSIDE THE DEVICE AND RESIDUAL FLUID AND CORROSION WAS FOUND ON THE SURFACES OF INTERNAL ASSEMBLIES. A LEAK TEST WAS PERFORMED, WHICH CONFIRMED THE PRESENCE OF A LEAK IN THE FLUID PATH CAUSED BY A TEAR IN THE CANNULA TUBING. THE LEAK WOULD HAVE RESULTED IN INSULIN COMING INTO CONTACT WITH INTERNAL COMPONENTS AND ASSEMBLIES, ULTIMATELY CAUSING THE DEVICE TO FAIL. A POD MALFUNCTION, THEREFORE, IS CONFIRMED TO HAVE BEEN A CONTRIBUTING FACTOR TO THE CUSTOMER'S HIGH BG LEVELS.

Description of Event or Problem · 1

THE REPORT INDICATED THAT THE CUSTOMER'S BG LEVELS WERE INCREASINGLY HIGH WHILE WEARING THIS POD (285-428 MG/DL); A HAZARD ALARM WAS INITIATED 14 HRS AFTER IT WAS ACTIVATED. AFTER REMOVING THE POD, HE TOOK IT APART AND NOTICED "THAT ONE OF THE BATTERIES LOOKED CORRODED" AND THAT "THERE WAS SOME BROWN LIQUID IN THE POD." THE DEVICE WILL BE RETURNED FOR EVAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L30393

Patients

Seq Age Sex Outcome Treatment
1 57 YR Other