FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1972614 · Received January 13, 2011

Report

Report Number
2531779-2011-00310
Event Type
Malfunction
Date Received
January 13, 2011
Report Date
December 16, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE INTACT WITHOUT ANY OBSERVATION OF PHYSICAL DAMAGE SUCH AS LIFTING OR PEELING. THE ISSUE WITH THE KEYPAD BUTTON MALFUNCTION WAS CONFIRMED THROUGH PRODUCT ANALYSIS. THE "UP, DOWN, AND OK" BUTTONS ARE NON-RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD MALFUNCTION. KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THERE WAS NO REPORT OF PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 19 YR