Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED, THE UP/DOWN/OK/CONTRAST KEYPAD BUTTONS INTERMITTENTLY RESPONDED TO USER INPUTS DURING TESTING, IT WAS OBSERVED THAT THE UP/DOWN/OK/CONTRAST KEYPAD CONTACTS BECAME INVERTED WHEN PRESSED AND DID NOT ALWAYS SPRING BACK, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE DOWN ARROW KEYPAD BUTTON REPORTEDLY NEEDS TO BE PRESSED SEVERAL TIMES FOR A RESPONSE. THE PT CLAIMED THAT THE BUTTON FEELS LOOSER THAN THE OTHER KEYPAD BUTTONS; HOWEVER, REPORTED NO VISIBLE DAMAGE TO THE KEYPAD. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.