FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1972591 · Received January 13, 2011

Report

Report Number
2531779-2011-00250
Event Type
Malfunction
Date Received
January 13, 2011
Report Date
December 15, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND WAS INVESTIGATED. THE PRIMING PROCESS WAS TESTED AND THE PROCESS PERFORMED SUCCESSFULLY. THE PUMP WAS PRIMED UNTIL THE PUMP INDICATED THAT 0 UNITS REMAINED IN THE CARTRIDGE; HOWEVER, WHEN THE CARTRIDGE WAS REMOVED 5 UNITS WERE VISUALLY FOUND REMAINING IN THE CARTRIDGE. EVAL REVEALED THAT THE PISTON CAP WAS MISSING, WHICH CONTRIBUTED TO THE INACCURATE INSULIN REMAINING VALUE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP WAS UNDER ESTIMATING THE ACTUAL AMOUNT OF INSULIN IN THE CARTRIDGE. THE PT DESCRIBED AN INCIDENT WHERE THE PUMP INDICATED THAT THERE WERE 147 UNITS OF INSULIN LEFT IN THE CARTRIDGE; HOWEVER, WHEN VISUALLY INSPECTING THE CARTRIDGE, IT HAS 160 UNITS. THE PUMP WAS REPLACED. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 46 YR