FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1972533 · Received December 30, 2010

Report

Report Number
3004209178-2010-83709
Event Type
Malfunction
Date Received
December 30, 2010
Date of Event
December 2, 2010
Report Date
December 2, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS EXPERIENCING LOW BLOOD GLUCOSE FOR A WEEK. THE MOTHER STATED THAT THE BASAL RATES ON THE INSULIN PUMP WERE REDUCED. TROUBLESHOOTING WAS PERFORMED. THE BOLUS HISTORY REVEALED TWO BOLUSES, WHICH THE CUSTOMER DID NOT DELIVER WHILE THE INSULIN PUMP WAS ON SUSPEND MODE. RAN A DISPLACEMENT TEST AND THE DEVICE PASSED THE TEST. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 12 YR