FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM

MDR report key: 19724089 · Received July 11, 2024

Report

Report Number
3002601200-2024-00300
Event Type
Malfunction
Date Received
July 11, 2024
Date of Event
June 5, 2024
Report Date
August 2, 2024
Manufacturer
BD SUZHOU (MDS)
Product Code
FOZ
UDI-DI
00382903830336
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

1. THE CUSTOMER RETURNED 2 PHOTOS, NO DEFECTIVE SAMPLE. THE PHOTOS SHOW: LOT NUMBER 4052074, LEAKAGE AT SMALL END OF CATHETER HUB. 2. DHR/BHR REVIEW LOT-4052074: 1-THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE 4 IN MARCH 2024, AND PACKAGED AT CFS PACKAGE LINE IN MARCH 2024. WORK ORDER QUANTITY WAS 198,000 EA. 2-REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. 3-REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 3. THE RETAINED SAMPLE OF THIS BATCH IS TAKEN FOR 45PSI LEAKAGE TEST, AND NO LEAKAGE IS FOUND AT THE CATHETER HUB. PLEASE SEE THE ATTACHED TEST REPORT. 4. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. CONCLUSION(S): THE RETURNED PHOTOS SHOW LEAKAGE AT THE SMALL END OF THE CATHETER HUB, AND THE ROOT CAUSE OF THIS DEFECT CANNOT BE DETERMINED BECAUSE NO ABNORMALITY IS FOUND ON PROCESS AND RETAINED SAMPLE, NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS, AND NO DEFECTIVE SAMPLE HAS BEEN RECEIVED FOR FURTHER TESTING. THE PLANT WILL CONTINUE TO TRACK AND TREND FOR THIS ISSUE.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMAITON PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM LEAKED AFTER APPLYING AN INDWELLING NEEDLE TO THE PATIENT, IT WAS FOUND THAT THERE WAS A LEAK OF LIQUID FROM THE NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236141 BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM INTRAVASCULAR CATHETER FOZ BD SUZHOU (MDS) 4052074 00382903830336

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown