FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT XCELERANT - HYDRO

MDR report key: 1972379 · Received January 25, 2011

Report

Report Number
2953200-2011-00164
Event Type
Injury
Date Received
January 25, 2011
Date of Event
December 26, 2010
Report Date
December 26, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070027
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: ENDOLEAK. DISEASE PROGRESSION WITH AORTIC NECK DILATATION. CONCLUSION: DISEASE PROGRESSION WITH AORTIC NECK DILATATION.

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5.2 CM ABDOMINAL AORTIC ANEURYSM APPROX 21 MONTHS AGO. THERE WAS MODERATE TORTUOSITY, MILD CALCIFICATION; AORTIC NECK WAS SHORT IN LENGTH (1CM) AND DISEASE PROGRESSION WITH AORTIC NECK DILATATION. AT A ROUTINE FOLLOW- UP APPOINTMENT, THE CT DEMONSTRATED DISTAL STENT GRAFT MIGRATION OF 8 MM, WITH A TYPE I ENDOLEAK. THE PHYSICIAN IMPLANTED A TALENT CUFF; HOWEVER, THE ENDOLEAK DID NOT RESOLVE (MFR REPORT # 2953200-2011-00165). THE PHYSICIAN IMPLANTED ANOTHER MANUFACTURER'S STENT GRAFT AND THE ENDOLEAK RESOLVED. NO CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT XCELERANT - HYDRO MIH MEDTRONIC CARDIOVASCULAR NA V00189895

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention