DEXTRUS 4137
Report
- Report Number
- 1028232-2011-00111
- Event Type
- Injury
- Date Received
- January 21, 2011
- Date of Event
- December 3, 2010
- Report Date
- January 17, 2011
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVZ
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEAD EXHIBITED LOSS OF CAPTURE AT 6.5 VOLTS AND IT WAS FOUND TO BE DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED WITH A TINED LEAD. TO DATE, THERE HAVE BEEN NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. AS OF THIS DATE, THE LEAD HAS NOT BEEN RETURNED. IF THIS LEAD SHOULD BE RETURNED TO OUR COMPANY, IT WILL BE ANALYZED AND THIS REPORT WILL BE REOPENED AND UPDATED AS NECESSARY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEXTRUS 4137 | PACER LEAD | NVZ | BIOTRONIK SE & CO. KG | 358764 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |