FDA Adverse Event
Injury
Summary report: N
PITT EASY ABUTMENT
MDR report key: 1972098
·
Received January 27, 2011
Report
- Report Number
- 9611820-2011-00003
- Event Type
- Injury
- Date Received
- January 27, 2011
- Report Date
- December 29, 2010
- Manufacturer
- SYBRON IMPLANT SOLUTIONS GMBH
- Product Code
- NHA
- PMA / PMN Number
- K053242
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
AN EVALUATION OF THE RETURNED SCREW INDICATED THAT THERE WAS NO DEFECTIVE MATERIAL. A VISUAL INSPECTION INDICATED THAT THE SCREW FRACTURED AT THE SCREW HEAD. THIS TYPE OF FRACTURE IS POSSIBLY DUE TO THE SUPERSTRUCTURE BEING UNDER CONSTANT TENSION OR THE SCREW BEING TIGHTENED WITH UNCONTROLLED TORQUE. THIS IS NOT A PRODUCT FAILURE. PLEASE NOTE THAT THIS REPORT IS BEING SUBMITTED BY SYBRON IMPLANT SOLUTIONS (B)(4) (THE MANUFACTURER) ON BEHALF OF SYBRON IMPLANT SOLUTIONS (THE IMPORTER) PER REPORTING EXEMPTION (B)(4).
Description of Event or Problem · 1
ON (B)(6), 2010 A DOCTOR REPORTED TO SYBRON IMPLANT SOLUTIONS (B)(4) THAT FIVE (5) PITT EASY ABUTMENT SCREWS FRACTURED. THIS MDR IS THE SECOND OF FIVE REPORTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PITT EASY ABUTMENT | ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS | NHA | SYBRON IMPLANT SOLUTIONS GMBH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other| R |