OT PING METER
Report
- Report Number
- 2939301-2011-00919
- Event Type
- Malfunction
- Date Received
- January 27, 2011
- Date of Event
- January 17, 2011
- Report Date
- January 18, 2011
- Manufacturer
- LIFESCAN INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K)# IS K082590.
FOLLOW-UP # 1 ((B)(4) 2012)-DEVICE EVALUATION: THE LAY USER/PATIENT'S PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ((B)(4) 2011) WITH THE FOLLOWING FINDINGS: THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE SPC PIN'S LIFTED HIGH. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE ERROR 5 METER ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN INC. | 3069610 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |