FDA Adverse Event Injury Summary report: N

PITT EASY ABUTMENT

MDR report key: 1972079 · Received January 27, 2011

Report

Report Number
9611820-2011-00005
Event Type
Injury
Date Received
January 27, 2011
Report Date
December 29, 2010
Manufacturer
SYBRON IMPLANT SOLUTIONS GMBH
Product Code
NHA
PMA / PMN Number
K053242
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION OF THE RETURNED SCREW INDICATED THAT THERE WAS NO DEFECTIVE MATERIAL. A VISUAL INSPECTION INDICATED THAT THE SCREW FRACTURED AT THE SCREW HEAD. THIS TYPE OF FRACTURE IS POSSIBLY DUE TO THE SUPERSTRUCTURE BEING UNDER CONSTANT TENSION OR THE SCREW BEING TIGHTENED WITH UNCONTROLLED TORQUE. THIS IS NOT A PRODUCT FAILURE. PLEASE NOTE THAT THIS REPORT IS BEING SUBMITTED BY SYBRON IMPLANT SOLUTIONS (B)(4) (THE MANUFACTURER) ON BEHALF OF SYBRON IMPLANT SOLUTIONS (THE IMPORTER) PER REPORTING EXEMPTION (B)(4).

Description of Event or Problem · 1

ON (B)(6), 2010 A DOCTOR REPORTED TO SYBRON IMPLANT SOLUTIONS GMBH THAT FIVE (5) PITT EASY ABUTMENT SCREWS FRACTURED. THIS MDR IS THE FOURTH OF FIVE REPORTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PITT EASY ABUTMENT ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS NHA SYBRON IMPLANT SOLUTIONS GMBH

Patients

Seq Age Sex Outcome Treatment
1 Other| R