FDA Adverse Event Malfunction Summary report: N

FLUSHING PUMP OFP-2 (EU)

MDR report key: 19719560 · Received July 11, 2024

Report

Report Number
9611174-2024-01316
Event Type
Malfunction
Date Received
July 11, 2024
Date of Event
June 11, 2024
Report Date
September 10, 2024
Manufacturer
KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
Product Code
FEQ
PMA / PMN Number
K100899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

E2/E3: NO INFORMATION AVAILABLE FOR THE OCCUPATION OF THE INITIAL REPORTER OR WHETHER THEY ARE A HEALTHCARE PROFESSIONAL THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE. THREE ATTEMPTS WERE PERFORMED TO OBTAIN ADDITIONAL INFORMATION, BUT NO RESPONSE WAS RECEIVED FROM THE CUSTOMER.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION AND A CORRECTION TO THE FOLLOWING FIELD(S): D8, D10, G3, H4, H6 AND H11. BASED ON THE RESULTS OF THE INVESTIGATION, AND SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION, THE MOST PROBABLE ROOT CAUSE OF THE REPORTED ISSUE IS EXPECTED OR RANDOM COMPONENT FAILURE WITHOUT ANY DESIGN OR MANUFACTURING ISSUE. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE FLUSHING PUMP HAD FAILURE AT THE PUMP HEAD. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE FLUSHING PUMP HAD FAILURE AT THE PUMP HEAD. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235857 FLUSHING PUMP OFP-2 (EU) OLYMPUS FLUSHING PUMP FEQ KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD. K10001143

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown