MDR report key: 19718796 · Received July 11, 2024

Report

Report Number
2028159-2024-01005
Event Type
Injury
Date Received
July 11, 2024
Date of Event
June 4, 2024
Report Date
October 15, 2024
Manufacturer
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
Product Code
OOE
UDI-DI
00380659981623
PMA / PMN Number
K163551
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION PROVIDED IN D.9., H.3., H.6., AND H.11. A MANUFACTURING DEVICE HISTORY RECORD REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. THE COMPANY REPRESENTATIVE WAS UNABLE TO CONFIRM NOR REPLICATE THE REPORTED EVENT. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. BASED ON THE INFORMATION OBTAINED, THE ROOT CAUSE OF THE REPORTED EVENT IS INCONCLUSIVE. A ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. BASED ON THE INFORMATION OBTAINED, THE ROOT CAUSE OF THE REPORTED EVENT IS INCONCLUSIVE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

H.3., H.6.: INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A CUSTOMER REPORTED AN ISSUE WITH A CAPSULE TEAR IN A PATIENT'S LEFT EYE DURING CATARACT SURGERY. THERE ARE MULTIPLE RELATED REPORTS FOR THIS FACILITY. THIS SPECIFIC REPORT ADDRESSES PATIENT'S DIJ WITH THE LEFT EYE ISSUE, AND ADDITIONAL REPORTS FROM OTHER MANUFACTURERS WILL BE FILED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235810 LENSX LASER SYSTEM OPHTHALMIC FEMTOSECOND LASER OOE ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER 550 14A0X3 00380659981623

Patients

Seq Age Sex Outcome Treatment
1 72 YR Female Other