SYNCHRON LX® I 725 CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-00253
- Event Type
- Malfunction
- Date Received
- January 27, 2011
- Date of Event
- December 28, 2010
- Report Date
- December 28, 2010
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K023049
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
QC PRIOR TO THE EVENT WAS SUCCESSFUL. QC AFTER THE EVENT FAILED WITH ELEVATED RESULTS ON LEVEL 3. GLUCOSE RE-CALIBRATION AFTER THE EVENT FAILED. THE LAST SUCCESSFUL CALIBRATION WAS ON (B)(4) 2010. BCI FIELD SERVICE ENGINEER (FSE) REPLACED THE REAGENT WITH ANOTHER CARTRIDGE OF THE SAME LOT AND ALL QC AND PRECISION TEST MET THE ESTABLISHED SPECIFICATIONS. THE ROOT CAUSE FOR THIS EVENT APPEARS TO BE THE REAGENT CARTRIDGE.
A CUSTOMER CONTACTED BECKMAN COULTER, INC. (BCI) IN REGARDS TO ERRONEOUSLY LOW GLUCOSE (GLU) RESULTS GENERATED BY SYNCHRON LX I 725 CLINICAL SYSTEM FOR TWO PATIENTS WHILE RUNNING IN DUPLICATES. THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. REPEAT TESTS ON THE SAME AND ON AN ALTERNATE INSTRUMENT PRODUCED HIGHER RESULTS. THERE WAS NO EFFECT TO THE PATIENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON LX® I 725 CLINICAL SYSTEM | CLINICAL CHEMSTRY ANALYZER | JJE | BECKMAN COULTER, INC. | LXI 725 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |