COBAS 6000 C501MODULE
Report
- Report Number
- 1823260-2011-00448
- Event Type
- Malfunction
- Date Received
- January 27, 2011
- Date of Event
- January 11, 2011
- Report Date
- March 30, 2011
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JFY
- PMA / PMN Number
- K060373
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SE
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
A SPECIFIC ROOT CAUSE WAS NOT IDENTIFIED. NO FURTHER ISSUES WERE REPORTED AFTER THE MAINTENANCE PERFORMED BY THE FIELD SERVICE REPRESENTATIVE.
THIS EVENT OCCURRED IN (B)(6).
THE USER SPORADICALLY RECEIVED QUESTIONABLE CREATININE PLUS GENERATION 2 (CREATININE) RESULTS OVER ONE AND A HALF MONTHS. FOUR PATIENT SAMPLES WERE INVOLVED; HOWEVER THE USER PROVIDED DATA FOR ONLY TWO PATIENT SAMPLES, OF WHICH ONLY ONE WAS REPORTED OUTSIDE THE LABORATORY. THE INITIAL RESULT WAS 609 UMOL/L AND WAS REPORTED OUTSIDE THE LABORATORY. THE PATIENT WAS ADMITTED TO THE ER FOR A "COUPLE OF HOURS" FOR EXAMINATION. THE USER STATED THIS "CAUSED STRESS ON BEHALF OF THE PATIENT". THE SAMPLE WAS REPEATED AND THE RESULT WAS 52.8 UMOL/L. THE RESULT FROM A NEW SAMPLE DRAWN FROM THE PATIENT A COUPLE OF HOURS LATER WAS 49.4 UMOL/L. THE PATIENT WAS NOT ADVERSELY AFFECTED AND WAS CURRENTLY FEELING FINE. THE CREATININE REAGENT LOT NUMBER WAS 630928. THE FIELD SERVICE REPRESENTATIVE FOUND A DROP OF ACID WASH ON THE WASHING STATION WHICH WOULD DRIP BACK INTO SOME OF THE CUVETTES. HE RESOLVED THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS 6000 C501MODULE | CLINICAL CHEMISTRY ANALYZER | JFY | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 051 YR | ALENNAT |