MENTOR SMOOTH ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2024-08196
- Event Type
- Injury
- Date Received
- July 10, 2024
- Date of Event
- June 24, 2024
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION REVIEW COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: BREAST PROSTHESIS DEFLATION. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
CORRECTION: IT WAS INCORRECTLY REPORTED THAT THE SUSPECT MEDICAL DEVICE IS A MENTOR SMOOTH ROUND MODERATE PROFILE 275CC SALINE BREAST PROSTHESIS. MENTOR RECEIVED ADDITIONAL EVENT INFORMATION ON 08-JUL-2024 THAT STATED THAT THE SUSPECT MEDICAL DEVICE IS A MENTOR SMOOTH ROUND MODERATE PROFILE 525CC SALINE BREAST PROSTHESIS, LOT #5687236, UDI #(B)(4). PMA #P990075. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. D4: UDI: AS THE SERIAL NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. ON 30-JUL-2024, MENTOR RECEIVED ADDITIONAL INFORMATION ABOUT THE EVENT. AN UPDATED IMPLANTATION DATE (B)(6) 2013 WAS REPORTED. THE REPORTED IMPLANTATION DATE IS AFTER THE EXPIRATION DATE OF THE SUSPECT MEDICAL DEVICE. ON 01-AUG-2024, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE ON THE SAME DATE. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, THE PATIENT EXPERIENCED DEFLATION IN THE BREAST IMPLANT. THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION AND LEAK TESTING OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE BREAST IMPLANT HAD A CREASE/FOLD EXTENDED FROM THE ANTERIOR TO THE POSTERIOR VIEW. LEAK TESTING WAS PERFORMED, ACCORDING TO THE MENTOR PROCEDURE, AND IT IDENTIFIED A TEAR WITHIN THE CREASE/FOLD ON THE POSTERIOR VIEW, MEASURING APPROXIMATELY 0.5 CM. THE EVALUATION DETERMINED THAT THE POSSIBLE CAUSE OF THE RUPTURE WAS CONSISTENT WITH NORMAL WEAR. DEFLATION CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. CREATING WRINKLES OR FOLDS SHOULD BE AVOIDED DURING IMPLANTATION OR OTHER PROCEDURES AS IT CAN RESULT IN DEFLATION AT A LATER TIME. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. IT WAS NOTED THAT, BASED ON THE DATE OF IMPLANTATION PROVIDED AND THE EXPIRATION DATE OF THE REPORTED LOT NUMBER, THE DEVICE WAS IMPLANTED AFTER THE EXPIRATION DATE. ACCORDING TO THE PRODUCT INSERT DATA SHEET, STERILITY CANNOT BE GUARANTEED IF THE PRODUCT IS USED AFTER THE ¿USE BY DATE¿ SHOWN ON THE BOX LABEL. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION PROCEDURE WITH A MENTOR SMOOTH ROUND MODERATE PROFILE 275CC SALINE BREAST PROSTHESIS THAT DEFLATED POST IMPLANTATION. DEFLATION OF THE PATIENT¿S RIGHT BREAST PROSTHESIS WAS DIAGNOSED VIA A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR SMOOTH ROUND MODERATE PLUS PROFILE 425CC SALINE BREAST PROSTHESES ON (B)(6) 2024.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 366968 | MENTOR SMOOTH ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 5687236 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Female | Required Intervention |