FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 1971279
·
Received January 26, 2011
Report
- Report Number
- 2050012-2011-00263
- Event Type
- Malfunction
- Date Received
- January 26, 2011
- Date of Event
- December 29, 2010
- Report Date
- December 29, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER CLEANED UP THE FLUID, BUT WAS UNABLE TO LOCATE THE CAUSE OF THE LEAK. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2010 AND FOUND THAT THE SAMPLE PROBE HAD GEL IN THE TUBING. THE FSE THEN VERIFIED THAT THE MODULE CHEMISTRY (MC) AND THE CARTRIDGE CHEMISTRY (CC) PROBES WERE CLEAR AND VERIFIED LEVEL SENSE ON THE SYSTEM.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT FLUID WAS LEAKING FROM THE REAGENT PROBE ONTO THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER COVERS. THERE WAS NO AFFECT ON PATIENT RESULTS OR OPERATOR EXPOSURE PERTAINING TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |