FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER.

MDR report key: 1971276 · Received January 26, 2011

Report

Report Number
2050012-2011-00214
Event Type
Malfunction
Date Received
January 26, 2011
Date of Event
December 31, 2010
Report Date
December 31, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QC RESULTS WERE WITHIN RANGE PRIOR TO AND POST THE EVENT. A BCI FIELD SERVICE ENGINEER (FSE) REPLACED THE GLUCOSE STIRRER MOTOR, REAGENT TUBING LINES AND REAGENT SYRINGE. ALTHOUGH HARDWARE WAS ADDRESSED, A CLEAR ROOT CAUSE CAN NOT BE DETERMINED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT A DIFFERENCE IN GLUCOSE MODULAR DUPLICATE RESULT ON THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER UNITS. THE CUSTOMER RUNS GLUM PATIENTS IN DUPLICATE ON THE INSTRUMENT AND THEY ARE NOT MATCHING. THE RESULT WAS NOT REPORTED OUT OF LABORATORY. PATIENT TREATMENT WAS NOT IMPACTED BY THIS EVENT.