UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER.
MDR report key: 1971276
·
Received January 26, 2011
Report
- Report Number
- 2050012-2011-00214
- Event Type
- Malfunction
- Date Received
- January 26, 2011
- Date of Event
- December 31, 2010
- Report Date
- December 31, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
QC RESULTS WERE WITHIN RANGE PRIOR TO AND POST THE EVENT. A BCI FIELD SERVICE ENGINEER (FSE) REPLACED THE GLUCOSE STIRRER MOTOR, REAGENT TUBING LINES AND REAGENT SYRINGE. ALTHOUGH HARDWARE WAS ADDRESSED, A CLEAR ROOT CAUSE CAN NOT BE DETERMINED.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT A DIFFERENCE IN GLUCOSE MODULAR DUPLICATE RESULT ON THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER UNITS. THE CUSTOMER RUNS GLUM PATIENTS IN DUPLICATE ON THE INSTRUMENT AND THEY ARE NOT MATCHING. THE RESULT WAS NOT REPORTED OUT OF LABORATORY. PATIENT TREATMENT WAS NOT IMPACTED BY THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER. | CLINICAL CHEMISTRY ANALYZER, | JJE | BECKMAN COULTER INC. | DXC800 PRO | N/A |