FDA Adverse Event
Malfunction
Summary report: N
IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM
MDR report key: 1971259
·
Received January 26, 2011
Report
- Report Number
- 2050012-2011-00261
- Event Type
- Malfunction
- Date Received
- January 26, 2011
- Date of Event
- December 29, 2010
- Report Date
- December 29, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K962294
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2010 AND REPLACED THE FILTER, FLUSHED THE TUBING AND PROBE WASH. A CHIP OF PAINT WAS FOUND IN THE PROBE WASH AND THIS ISSUE HAS BEEN RESOLVED. CUSTOMER RAN QC AND PATIENT SAMPLES WITH NO ERRORS.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE SAMPLE PROBE WAS FLOODING IN THE IMMAGE 800 IMMUNOCHEMISTRY SYSTEM. NO OPERATOR EXPOSURE WAS NOTED FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | IMMAGE 800 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |