FDA Adverse Event Malfunction Summary report: N

IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM

MDR report key: 1971259 · Received January 26, 2011

Report

Report Number
2050012-2011-00261
Event Type
Malfunction
Date Received
January 26, 2011
Date of Event
December 29, 2010
Report Date
December 29, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K962294
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2010 AND REPLACED THE FILTER, FLUSHED THE TUBING AND PROBE WASH. A CHIP OF PAINT WAS FOUND IN THE PROBE WASH AND THIS ISSUE HAS BEEN RESOLVED. CUSTOMER RAN QC AND PATIENT SAMPLES WITH NO ERRORS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE SAMPLE PROBE WAS FLOODING IN THE IMMAGE 800 IMMUNOCHEMISTRY SYSTEM. NO OPERATOR EXPOSURE WAS NOTED FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. IMMAGE 800 N/A

Patients

Seq Age Sex Outcome Treatment
1