FDA Adverse Event Injury Summary report: N

CAPSURE EPI

MDR report key: 1971236 · Received January 26, 2011

Report

Report Number
2182208-2011-00029
Event Type
Injury
Date Received
January 26, 2011
Date of Event
January 19, 2011
Manufacturer
MEDTRONIC, INC.
Product Code
DTB
PMA / PMN Number
P950024/S2
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Additional Manufacturer Narrative · 1

THE LEAD HAS BEEN RETURNED TO THE MANUFACTURER. THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND THE DISTAL CONDUCTOR FRACTURED. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THE LEAD WAS STRETCHED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD IMPEDANCE WAS OUT OF RANGE, THERE WAS A LEAD INTEGRITY ALERT, NO VENTRICULAR PACING AND THERE WAS A LEAD FRACTURE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD IMPEDANCE WAS OUT OF RANGE, THERE WAS A LEAD INTEGRITY ALERT, NO VENTRICULAR PACING AND THERE WAS A LEAD FRACTURE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD IMPEDANCE WAS OUT OF RANGE, THERE WAS A LEAD INTEGRITY ALERT, NO VENTRICULAR PACING AND THERE WAS A LEAD FRACTURE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.IT WAS REPORTED THAT THE LEAD IMPEDANCE WAS OUT OF RANGE, THERE WAS A LEAD INTEGRITY ALERT, NO VENTRICULAR PACING AND THERE WAS A LEAD FRACTURE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. (B)(6) 2011, THE LEAD HAS BEEN RETURNED TO THE MANUFACTURER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE EPI IMPLANTABLE PACING LEAD DTB MEDTRONIC, INC. 4968 ASKU

Patients

Seq Age Sex Outcome Treatment
1 1 YR Hospitalization| O| R (B)(4) IMPLANTABLE PULSE GENERATOR| 4968 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR| (B)(4) IMPLANTABLE PULSE GENERATOR| 4968 IMPLANTABLE PACING LEAD| 4968 IMPLANTABLE PACING LEAD