CAPSURE EPI
Report
- Report Number
- 2182208-2011-00029
- Event Type
- Injury
- Date Received
- January 26, 2011
- Date of Event
- January 19, 2011
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DTB
- PMA / PMN Number
- P950024/S2
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
THE LEAD HAS BEEN RETURNED TO THE MANUFACTURER. THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND THE DISTAL CONDUCTOR FRACTURED. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THE LEAD WAS STRETCHED.
IT WAS REPORTED THAT THE LEAD IMPEDANCE WAS OUT OF RANGE, THERE WAS A LEAD INTEGRITY ALERT, NO VENTRICULAR PACING AND THERE WAS A LEAD FRACTURE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE LEAD IMPEDANCE WAS OUT OF RANGE, THERE WAS A LEAD INTEGRITY ALERT, NO VENTRICULAR PACING AND THERE WAS A LEAD FRACTURE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE LEAD IMPEDANCE WAS OUT OF RANGE, THERE WAS A LEAD INTEGRITY ALERT, NO VENTRICULAR PACING AND THERE WAS A LEAD FRACTURE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.IT WAS REPORTED THAT THE LEAD IMPEDANCE WAS OUT OF RANGE, THERE WAS A LEAD INTEGRITY ALERT, NO VENTRICULAR PACING AND THERE WAS A LEAD FRACTURE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. (B)(6) 2011, THE LEAD HAS BEEN RETURNED TO THE MANUFACTURER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE EPI | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC, INC. | 4968 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 1 YR | Hospitalization| O| R | (B)(4) IMPLANTABLE PULSE GENERATOR| 4968 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR| (B)(4) IMPLANTABLE PULSE GENERATOR| 4968 IMPLANTABLE PACING LEAD| 4968 IMPLANTABLE PACING LEAD |