FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 1971218 · Received January 26, 2011

Report

Report Number
2031642-2011-00015
Event Type
Malfunction
Date Received
January 26, 2011
Date of Event
December 29, 2010
Report Date
December 29, 2010
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

POWER SUPPLY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE VENTILATOR WOULD NOT OPERATE ON AC POWER, ONLY ON BACKUP BATTERY POWER. THE MANUFACTURER'S SERVICE TECHNICIAN WAS ABLE TO DUPLICATE THE REPORTED PROBLEM. THE SERVICE TECHNICIAN REPLACED THE POWER SUPPLY TO CORRECT THE REPORTED PROBLEM. APPLICABLE FINAL TESTING WAS PERFORMED AND TESTS PASSED TO OPERATING SPECIFICATIONS. IF A FAILURE OF THE POWER SUPPLY OCCURS DURING USE AND THE VENTILATOR SHUTS DOWN DUE TO A LOSS OF AC POWER, AN AUDIBLE POWER FAIL ALARM WILL ACTIVATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA