WALLFLEX ENTERAL DUODENAL STENT WITH ANCHOR LOCK DELIVERY SYSTEM, MOD M00565010
Report
- Report Number
- 3005099803-2011-00184
- Event Type
- Malfunction
- Date Received
- January 26, 2011
- Date of Event
- January 4, 2011
- Report Date
- January 4, 2011
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- MUM
- PMA / PMN Number
- K062750
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KR
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE COMPLAINANT INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. BASED ON THE INFORMATION AT HAND, THE MOST PROBABLE ROOT CAUSE IS BEING LABELED AS OPERATIONAL CONTEXT. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT. A LABELING REVIEW WAS PERFORMED AND NO DEVIATION WAS FOUND.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A WALLFLEX ENTERAL DUODENAL STENT WAS USED DURING A STENTING PROCEDURE WITHIN THE SIGMOID COLON ON (B)(6), 2011. ACCORDING TO THE COMPLAINANT, THE PHYSICIAN COULD NOT COMPLETELY DEPLOY THE STENT; AFTER ABOUT HALF HAD BEEN DEPLOYED, IT COULD NOT BE FURTHER DEPLOYED NOR RECONSTRAINED. THE PHYSICIAN REMOVED THE DEVICE FROM THE PATIENT AND USED A DIFFERENT TYPE OF STENT TO COMPLETE THE PROCEDURE. THE PHYSICIAN ADDED THAT THE ANATOMY WAS VERY TORTUOUS, AND THAT THE STRICTURE WAS ASSOCIATED WITH A TUMOR. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WALLFLEX ENTERAL DUODENAL STENT WITH ANCHOR LOCK DELIVERY SYSTEM, MOD M00565010 | STENT,METALLIC,EXPANDABLE,DUODENAL | MUM | BOSTON SCIENTIFIC - GALWAY | M00565020 | 13389678 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |