FDA Adverse Event Malfunction Summary report: N

FREESTYLE FREEDOM LITE

MDR report key: 1971099 · Received January 26, 2011

Report

Report Number
2954323-2011-01074
Event Type
Malfunction
Date Received
January 26, 2011
Date of Event
December 29, 2010
Report Date
January 26, 2011
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. NOTE: THE METER IS DESIGNED TO DISPLAY READINGS OF 1.1 MMOL/L TO 27.8 MMOL/L. THIS METER DOES NOT SHOW A NUMERIC VALUE GREATER THAN 27.8 MMOL/L. A BLOOD GLUCOSE READING ABOVE 27.8 MMOL/L WILL READ AS A "HI" IN THE ADC METER.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 50 MMOL/L, 12.6 MMOL/L, AND 8.3 MMOL/L WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL OUT OF RANGE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE FREEDOM LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 1013218

Patients

Seq Age Sex Outcome Treatment
1