LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM (UNK SIZE)
Report
- Report Number
- 2024601-2010-01058
- Event Type
- Malfunction
- Date Received
- January 5, 2011
- Date of Event
- April 15, 2009
- Report Date
- December 14, 2010
- Manufacturer
- ALLERGAN
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
TAPER UNK. MEDWATCH SENT TO FDA ON: (B)(4) 2011. THE REPORTER OF THE COMPLAINT WAS ASKED TO RETURN THE PRODUCT FOR ANALYSIS AS WELL AS TO INDICATE THE PRODUCT SERIAL NUMBER. THE INFO HAS NOT YET BEEN RECEIVED BY ALLERGAN. THE CONNECTOR TYPE CANNOT BE IDENTIFIED NOR AN ASSUMPTION BE MADE AS TO THE TYPE OF CONNECTOR ASSOCIATED WITH THIS COMPLAINT BECAUSE NO SERIAL NUMBER WAS GIVEN. VISUAL EXAMINATION MAY DETERMINE THE CONNECTOR TYPE ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFO HAS BEEN REPORTED TO ALLERGAN REGARDING THE SERIAL NUMBER. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."
HEALTH PROFESSIONAL REPORTED THAT THE ACCESS PORT OF A LAP-BAND DEVICE WAS REPLACED BECAUSE IT WAS LEAKING. THE DOCTOR BELIEVED IT WAS A LEAK AFTER NUMEROUS VISITS SHOWED LESS FLUID THAN WHAT WAS INJECTED. THE HEALTH PROFESSIONAL GAVE AN EXAMPLE AS THE OFFICE "WOULD FILL THE DEVICE WITH 3.2 CC OF SALINE BUT WOULD ONLY BE ABLE TO REMOVE 2.8 CC ON EACH VISIT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM (UNK SIZE) | LTI | ALLERGAN | NA | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |