NV GXL LNR, +5LAT, 36MM G2-52/54MM CUPS
Report
- Report Number
- 1038671-2024-02337
- Event Type
- Injury
- Date Received
- July 10, 2024
- Date of Event
- August 29, 2023
- Report Date
- February 13, 2025
- Manufacturer
- EXACTECH, INC.
- Product Code
- JDI
- UDI-DI
- 10885862024329
- PMA / PMN Number
- K070479
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
BASED ON THE AVAILABLE INFORMATION, THE PATIENT MEETS THE FOLLOWING RISK CRITERIA FOR EARLY PROSTHESIS WEAR AND/OR OSTEOLYSIS: IMPLANTED WITH A LATERALIZED LINER, COMBINATION OF LARGEST AVAILABLE FEMORAL HEAD AND/OR THINNEST AVAILABLE ACETABULAR LINER WAS USED AND IMPLANTED WITH A COMPONENT HAVING A SHELF AGE OF GREATER THAN 2 YEARS. THE MOST LIKELY CAUSE FOR THE REPORTED REVISION DUE TO PROSTHESIS WEAR AND OSTEOLYSIS MAY HAVE BEEN A COMBINATION OF RISK FACTORS SUCH AS USE ERROR, IMPLANT POSITIONING, IMPLANT SIZE SELECTION, AND PATIENT FACTORS (FITNESS FOR SURGERY, BIOMECHANICS, ACTIVITY LEVEL AND LOCAL TISSUE OXIDATION POTENTIAL). HOWEVER, THIS CANNOT BE CONFIRMED AS THE DEVICES WERE NOT RETURNED FOR EVALUATION, AND IMAGES AND RADIOGRAPHS WERE NOT PROVIDED. H6: CORRECTED HEALTH EFFECT AND MEDICAL DEVICE CLINICAL CODES.
D10: CONCOMITANT DEVICES: 5207638, 190-30-04 - ALT HA S CLR STD SZ 4; 5414381, 170-36-93 - BIOLOX DELTA FEMORAL HEAD 36MM OD, -3.5MM; 5918747, 186-01-52 - INTEGRIP CC, CLUSTER 52MM, G2; 6201653, 101-05-20 - 3.2MM DRILL BIT 20MM 1PK; S038520, 180-65-20 - ALTEON 6.5MM SCREW, 20MM. THE PRODUCT ASSOCIATED WITH THE REPORTED EVENT IS WITHIN THE SCOPE OF RECALL Z-1729-2022; HOWEVER, THERE IS INSUFFICIENT INFORMATION TO EVALUATE WHETHER THE SUBJECT ISSUE OF THE RECALL WAS THE CAUSE OR CONTRIBUTOR TO THE REPORTED EVENT. THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO MEDICAL OR OTHER RECORDS CONTAINING TREATMENT INFORMATION OR PATIENT INFORMATION HAVE BEEN RECEIVED; THEREFORE, THE REPORTED EVENT CANNOT BE CONFIRMED, NOR CAN THE CIRCUMSTANCES OR POTENTIAL CAUSES OR CONTRIBUTORS TO THE ALLEGED EVENT BE EVALUATED. SHOULD ADDITIONAL MATERIAL INFORMATION BECOME AVAILABLE THAT PERMITS MORE ANALYSIS OR CONCLUSIONS, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
IT WAS REPORTED THAT APPROXIMATELY 46 MONTHS AFTER A LEFT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, LEFT HIP PAIN, SNAPPING, SCARRING, AND IRRITATION; EVIDENCE OF POLYETHYLENE IRRITATION; CRACKING AND ECCENTRIC WEAR OF POLYETHYLENE. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN SATISFACTORY CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1281641 | NV GXL LNR, +5LAT, 36MM G2-52/54MM CUPS | PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED | JDI | EXACTECH, INC. | 10885862024329 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention| H | SEE H11. |