AUTOSOFT XC
Report
- Report Number
- 3003442380-2024-13315
- Event Type
- Malfunction
- Date Received
- July 10, 2024
- Date of Event
- June 3, 2024
- Report Date
- November 25, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244016712
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. REVISION 21 OF RECORD (B)(4) DOES NOT REQUIRE A TYPE 2 REPORTABLE COMPLAINT TO INITIATE A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS CREATED AND SUBSEQUENTLY CLOSED TO DOCUMENT THE DEVICE HISTORY RECORD (DHR) REVIEW, WHICH WAS NECESSARY TO ADVANCE THE PARENT RECORD TO THE REVIEW AND SUMMARY. COMPLAINT INVESTIGATION: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6000578 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6000578 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 103 AND MANUFACTURED IN THE LINE # 3 ON 28/APR/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, ONE OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION, HARM CODE AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION OR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED EVENT OF INSULIN FLOW BLOCK AT THE SITE ON (B)(6) 2024. THE EVENT OCCURRED AFTER 3 OR MORE HOURS OF INSERTION. THE INSERTION SITE WAS AT THE ABDOMEN. THE BLOOD WAS OBSERVED AT THE SITE. NO FURTHER INFORMATION WAS AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 180615 | AUTOSOFT XC | UNO INSET I 60/6 BLUE TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1001683 | 6000578 | 05705244016712 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Female |